next up previous contents
Next: E-9: A.C. Circuits Up: Electricity and Magnetism Previous: EC-7d  Induction - Faraday Discovery   Contents


E-8: Cathode Ray Oscilloscope and Differential Amplifiers

OBJECTIVES: To learn the basic operation of an oscilloscope in order to observe the voltage vs time relationships of basic electrical signals.

Preliminary Question:

APPARATUS:

Dual-trace oscilloscope plus manual, signal generator plus frequency counter, digital multimeter, (DMM); circuit plug board & component kit; differential amplifier.

Part A - Operation of the Oscilloscope
(Leader, Model 1021, or Hitachi, Model V-202F)

INTRODUCTION:

The oscilloscope is basically a graph-displaying device - it draws a graph of an electrical signal. In most applications the graph shows how signals change over time: the vertical (Y) axis represents voltage and the horizontal (X) axis represents time. The intensity or brightness of the display is sometimes called the Z axis. This simple graph can tell you many things about a electrical signal. Here are just a few, you can ...

Your dual trace analog oscilloscopes have circuits which can alternately display sweep signals from two different input channels (Y$_1$ and Y$_2$). Unfortunately this instrument can address only a single point on the display window at one instant in time. However the screen is coated with a phosphor coating which whose image persists for a short time. When this property is combined with the finite time resolution of the eye one can observe ``simultaneous'' dual traces.

\fbox{{\em CAUTION:}
Avoid placing signal generator on top of scope.}
[Stray magnetic fields from power supplies can interfere with the cathode ray tube (i.e., the display or CRT) operation and create a noisy (fuzzy) ``trace.'']

If you are unfamiliar with the various control knobs and switches you should read through this section carefully.

SUGGESTED PROCEDURE (refer to Fig. 1):

  1. Using a BNC cable and the sinusoidal wave signal output from the function generator (1kHz, 1V rms) connect the signal to the CH1 input (# 7) oscilloscope (or scope). Note that each input, CH1 and CH2, has a three position switch (#6 or #15) for either grounding the input (GND), coupling the signal through a capacitor (AC), or a direct connection (DC). Using the AC input (use this position) removes all DC components of the input signal.
    Figure 1
    \includegraphics[height=7.2in]{figs/e8-01.eps}

  2. Turn on the scope (#1)and note the four main groups of controls:
    i.
    those associated with the intensity and focus of the trace
    ii.
    those controlling vertical deflection (#6 thru 15).
    iii.
    those controlling horizontal deflection (#16 thru 19).
    iv.
    those associated with initiating the trace movement, TRIGGER,(#20 thru 24).
  3. Nominal starting positions: (If you have trouble finding the beam consult your instructor.)
    LEVEL(20) 0 (or so) VMODE(10) CH1
    MODE(22) NORM VOLT/DIV 0.2 V
    COUPLING(23) DC RED knobs fully CW (to CAL)
    SOURCE(24) CH1 AC/GND/DC AC
    HOLDOFF(19) fully CCW POSITIONs (9,12,17) centered
    TIME VARIABLE(16) fully CW INTEN/FOCUS/ILLUM centered
    TIME/DIV(18) 0.2 ms 9,12 and 16 pushed in

Additional desciption of oscilloscope controls
A:
The INTENSITY knob controls the trace brightness3by varying the potential of a control grid or an aperture near the electron emitting cathode. The FOCUS adjusts the trace sharpness by varying the potential on the focusing anode. POSITION knobs move the traces up and down, and also right and left.
B:
The VOLTS/DIV and red VARIABLE knobs provide coarse and fine adjustment of the pattern height by varying the amplification of the signals from the input terminals. To read the correct volts/division shown on the scales, the red variable knobs must be in the CAL (calibrated) position (fully clockwise).

The V MODE lever permits display of either or both channels; for both use either ALT for alternate sweeps, or CHOP which switches channels at a 250 kHz rate. (Hitachi's DUAL automatically switches from ALT to CHOP at a TIME/DIV setting = 1 ms/DIV).

Except for the Hitachi, pulling out the CH1 POSITION control (PULL ADD) adds both channels (CH1 + CH2), and pulling out the CH2 POSITION control (PULL INV) inverts its signal (-CH2). Hence using both PULL ADD and PULL INV gives the difference of the two channels (CH1 + -CH2). (For the Hitachi use the ADD or DIFF on the vertical MODE lever).

C:
The TIME/DIV and VARIABLE or SWP/VAR knobs give coarse and fine control of the trace horizontal sweep motion. Again for the TIME/DIV scales to be correct, set the VARIABLE (or SWP/VAR) knob at the CAL position (fully clockwise).

Once the trigger requirements have been satisfied, the scope sweeps the trace horizontally from left to right at a uniform rate (set by the TIME/DIV knob), and then quickly returns the trace to the start position. These trigger circuits are useful for synchronizing the sweep start with some common feature of the input signal although the signal itself may not be perfectly periodic (e.g. your heartbeat). (see Fig. 2).

Figure 2
\includegraphics[height=3.8in]{figs/e8-02.eps}

D:
The TIME/DIV knob, when set to the \framebox{X-Y} position, disconnects the internal horizontal sweep generator and uses the CH1 X input to specify the horizontal sweep; CH2 is still connected for vertical displacements and thus becomes Y. (For Hitachi also set MODE lever to CH 2 X-Y). The resultant X-Y pattern forms a Lissajous figure.

E:
Other triggering options:

Part B - Suggested Oscilloscope Experiments

  1. Observe both sine and square wave signals of various frequencies. Experiment with setting to learn the effect of different controls, (e.g. automatic vs triggered sweeps). For the square waves look both at low frequencies and at maximum frequencies. Notice any effects of changing the input coupling (#6 or 15) (switch back and forth from AC to DC). Sketch the appearance of ``square'' waves at high and low frequencies. If possible, explain the differences.

    i.
    Verify the calibration of the scope's Y scaling by using the 0.5 volt peak to peak square wave signal from the calibration terminal (CAL or #11). To connect this signal to the scope, just touch the exposed calibration terminal with the center gold pin of the bayonet coaxial (BNC) connector on the input cable.

    ii.
    Use a digital multimeter (DMM) to measure both the sine wave and square wave voltage output of the signal generator. Avoid very high frequencies since capacitative loading may then be a problem when the signal generator is also connected to the scope.

    iii.
    Compare the DMM readings (which are r.m.s. or ``effective'' voltages) with the peak-to-peak voltages reading directly off the scope screen. Are they consistent?
    Remember the voltage amplitude is one-half the peak to peak voltage and the rms voltage is

    \begin{displaymath}\frac{\mbox{peak voltage}}{\sqrt{2}}=\frac{\mbox{peak-to-peak voltage}}{2.818}  !\end{displaymath}

  2. OPTIONAL: Check the calibration of the scope's X (horizontal) scaling by using the TIME/DIV reading to measure the period for one wavelength of the signal. Compare the deduced frequency with signal generator scale setting and/or that of a frequency meter.

OPTIONAL:

Lissajous Figures: When two periodic motions at right angles combine, the resulting pattern is generally very complex. However, sine waves whose frequencies are in small integer ratios give simple stationary patterns. For example see Fig. 3.

To produce Lissajous figures on the scope, set the TIME/DIV control to \framebox{X-Y}. Connect the 60 Hz terminal to the CH1 X input. To the CH2 Y input apply a sine wave input whose frequency bears a simple ratio with respect to 60 Hz. (For the Hitachi, also set the vertical MODE lever to CH 2, X-Y.)

A Lissajous pattern should result. Adjust amplitudes and positions to locate the figure properly on the screen. The frequency scale on the signal generator may not be accurate, but the 60 Hz power line is an excellent frequency standard.
Use Lissajous figures to calibrate your signal generator at 120 Hz and 180 Hz.

Observe and explain patterns with the frequency ratios of Fig. 3. Also try a ratio 2:3.

Figure 3: Examples of Lissajous figures.
\includegraphics[height=2.2in]{figs/e8-04.eps}

Part C - Differential Amplifiers
INTRODUCTION:
Amplifiers are devices which usually increase the amplitude of the output signal compared to the input signal. The customary symbol for an amplifier is a triangular shape:

Figure 4
\includegraphics[height=1.4in]{figs/e8-05.eps}

In the sine wave example above, the gain is 10 and the input signal is between an input terminal and ground. The internal oscilloscope amplifiers for channels Y$_1$ and Y$_2$ are of this type and have a common ground. Because of this common ground, one has problems in using a scope to examine simultaneously voltages across individual circuit elements that are in series (see below, also E9 lab).

We can avoid these problems by interposing ``differential amplifiers'' which have two inputs $V'$ and $V''$, (neither at ground), and which amplify only the voltage difference ($V'-V''$). Symbolically:

Figure 5
\includegraphics[height=1.5in]{figs/e8-06.eps}

Note now that the ground of the output signal is independent of any input ground.

SUGGESTED EXPERIMENTS:

1.
Become familiar with the circuit plug board (Fig. 6) and the banana type plug-in components. Note there are 24 groups with 9 socket plug-in holes in each group. While the 9 plug-in holes in each group are electrically connected together, each of the 24 groups is isolated unless connected by plug-in components. Two isolated metal bus bars (with banana plug-in holes) on the sides may facilitate common connections for some circuits.

Figure 6
\includegraphics[height=3.2in]{figs/l104/e08-2.eps}

The plug-in components (metal bridges, resistors, capacitors, inductors, or other circuit elements) will only connect between adjacent groups of the 9 (already connected) plug-in holes or from the edge metal bus bars to an adjacent 9 hole group. We also provide coaxial cable to banana plug-in connectors: these facilitate using signal generators, scopes, etc. with the circuit plug-in board. Be sure to notice the raised retangular protuberance on one side of the connector. This indicates the connector ground bar side.

\includegraphics[height=2.2in]{figs/l104/n02-3.eps}

As a first configuration use the RC series circuit sketched in Fig. 6 as a guide using the scope to observe simultaneously the output of the AC signal generator and the signal across the resistor. How do the relative $V_{\mbox{\tiny peak}}$-to- $V_{\mbox{\tiny peak}}$ amplitudes compare at low, medium and high frequencies. It may help to know that the reactance $X_C$ of a capacitor is $X_C = 1/\omega C$, the impedence $Z$ is $Z=\sqrt{R^2+X^2_C}$ and medium frequencies are given by $X_C \approx R$. Qualitatively explain the behavior.

Do to voltage peaks always occur at the same time?

2.
The phase shift $\phi$ between the current (monitored by the voltage drop across the resistor) and the voltage has the relation $\tan \phi =
-X_C/R=V_C/V_R$. To observe this behavior you must simultaneously monitor the signal across the resistor and capactors. Unfortunately, only one of these components does reference to ground; a necessary aspect of your scope inputs. (Which one?) To compensate for this short coming a module containing dual differential amplifiers is provided.

Connect the resistor and capacitor signal to the differential amplifiers as shown in the figures below. Connect the outputs of the differential amplifiers to the scope CH 1 and CH 2 inputs, and set both gains to 1.

\includegraphics[width=2.2in]{figs/l104/e08-7.eps} Figure 7: Dual differential amplifier module. \includegraphics[width=3.6in]{figs/l104/e08-8.eps} Figure 8: Inputs to differential amplifiers.

Sketch the oscilloscope display when $X_C=R$ and show how you can use this information to obtain the phase shift. Compare your result to that of the $\tan \phi$ formula.

3.
Replace the capacitor with the diode and use the differential amplifiers to now observe the action of a silicon diode on an A.C. signal.

NOTE: Silicon diodes (Fig. 9) are junctions between silicon semiconductors with different type doping: n-type with an impurity which easily donates electrons, e.g. phosphorus (P) in silicon;4 and p-type with an ``acceptor'' impurity, e.g. aluminium (Al) in silicon$^5$. The junction between the two types becomes a very non-linear device which if subjected to an electric field acts almost as a one way valve to current flow. The diode symbol has the arrow pointed in the direction of conventional easy current flow (from p-type to n-type) though actual electron flow may be in the opposite direction.

Figure 9
\includegraphics[height=1.7in]{figs/e8-08.eps}

SUGGESTED EXPERIMENT:

Set up the circuit plug board as shown in Fig. 10. Connect Y$_1$ and Y$_2$ to the two scope inputs.

Figure 10
\includegraphics[height=2.0in]{figs/e8-09.eps}

Vary the amplitude of the signal generator input and sketch the resultant waveforms across the resistor and across the diode.


next up previous contents
Next: E-9: A.C. Circuits Up: Electricity and Magnetism Previous: EC-7d  Induction - Faraday Discovery   Contents
Physics Laboratory 2001-08-28