Next: L-8: Polarization
Up: Light
Previous: L-3: Optical Instruments
  Contents
L-5: Spectrometer and the H Balmer Series
NOTE: Consult your instructor for assignment of
parts suitable for a single lab, e.g. do
Part I, II (A or B), Part IV or Part V).
OBJECTIVES:
- To become familiar with a precision spectrometer and some of its
uses.
APPARATUS:
- Precision spectrometer; mercury and hydrogen discharge tubes;
prism; diffraction grating; light source for Gaussian eyepiece
& its 12V power supply (L-2); ring stand & achromatic lens.
Figure 1:
The spectrometer (side view).
|
INTRODUCTION and ADJUSTMENTS
PART I - A
- Become familiar with the clamping and fine adjustment controls for
telescope and prism table angles. Never force a motion - you
may damage the instrument. If at the end of a fine adjustment,
loosen and reclamp! The prism table clamp sets table
elevation: no fine adjustment is necessary.
- Note that 180
apart are two angle
scale reading ports and verniers, I and J.
The verniers have 30 divisions per half degree. Hence each division is
(1/30)
(1/2) = (1/60)
or one minute (
) of arc.
- The knurled ring about the collimator, B, controls the slit opening.
- Staff have already focussed the collimator for parallel light and
have set collimator and telescope
to spectrometer's rotation axis.
REMINDER ON READING A VERNIER
PART I - B: FOCUS THE TELESCOPE FOR PARALLEL RAYS
- 1.
- i)
- Slide telescope eyepiece in or out until crosshairs are in sharp
- ii)
- Sight
telescope at a distant object (thru an open window); then focus
telescope by rotating its focus ring
(Fig. 1) until the object's clear image falls on the cross
hairs. The test for proper focus is absence of parallax
between the image and cross hairs (appendix 4). The telescope, now
focused for parallel rays, will stay so as long as the focus ring is
unmoved; but one may still adjust the eyepiece to suit the observer.
- 2.
- An alternative method to focus for parallel light is to use the
Gaussian eyepiece + light source as described below in PART II B.
- 3.
- With collimator and telescope both properly focused one
should get a sharp image of the slit and no parallax between slit image
and cross hairs. If you still get parallax, recheck your
telescope focus and/or consult instructor.
- 4.
- Optimizing light thru the collimator: A properly located short focal
length lens can gather a large fraction of the light from a source and
redirect it
thru the collimator thereby facilitating detection of weak
spectral lines.
For strong
lines it permits narrowing the slit width and thus improving resolution.
Use such a lens (mounted on a ring stand) to find an arrangement which
fills the collimator with light.
Figure 3:
Proper, a), and improper, b) prism locations.
|
PART II-A: MEASUREMENT OF PRISM ANGLE
- Mount the prism (in holder) on the dowel pins (in the prism
table) so that the prism is far enough from the collimator that beams B
and B
are centered with respect to the spectrometer axis; otherwise much
of the reflected light may miss the telescope. Turn the table so that
the apex angle A (Fig. 3) splits the
beam into nearly equal parts.
- Using the prism table levelling screws adjust the plane of the
prism table so it is close to horizontal.
- With the prism table clamped at the proper elevation, set
the telescope to receive beam B
and form an image of the slit on the intersection of the telescope cross hairs.
Record both VERNIER readings (in minutes).
Average the two vernier readings
(to eliminate any systematic error from misalignment of the circle
scale with respect to bearing axis),
and add the result to one of the angle scale readings. Note
that the angle
scale reads in half-degree units and the vernier in minutes (not in decimal
degrees!). The zero position on the vernier determines
the angle to the nearest
half-degree (30 minutes) and the vernier reads the number of minutes past the
half-degree mark.
- Then with telescope in the 2nd position,
set on the slit image. The change in reading of the
same angle scale should be the angle D. (See Fig. 3.) Be sure you don't
get the angle scales mixed up and subtract the first reading of angle scale
1 from a reading of angle scale 2. Also some students incorrectly handle
the subtraction when the scale passes through
or
.
- OPTIONAL: Try to prove that angle D is twice the prism angle A.
PART II-B: ALTERNATE method for prism angle measurement using Gaussian eyepiece
Figure 4:
Spectrometer telescope
|
- Introduction:
The Gaussian eyepiece (Fig. 4)
has a partially reflecting
glass plate G set at 45
to telescope axis so that
light from the lamp reflects down the telescope tube, past
the cross hairs and out the objective. If in front of the objective
you place a reflecting surface perpendicular (
) to the telescope axis,
the light will reflect back into the objective and form
a real image of the
cross hairs. If the cross hairs are in the focal
plane of the objective, their reflected image will form in the
same plane. Thus both the
cross hairs and the image will be in focus through the eyepiece.
When one orients the
reflecting surface so that the image coincides with the cross
hairs, the reflector is accurately
to telescope axis.
Thus the Gaussian eyepiece permits both focusing the telescope for
parallel rays and setting a reflecting surface
to the
telescope axis.
- 1.
- To focus the telescope for parallel light:
- a)
- Rotate the Gaussian eyepiece to open the hole between
reflecting plate G and lamp. Adjust eyepiece to
give a sharp image of the cross hairs, but don't turn the
eyepiece to block the light hole. Next, with prism on the table, adjust
the prism table screws to make
the table nearly horizontal.
- b)
- Now rotate the prism table until the prism
face is approximately
to telescope axis. Clamp the telescope and mount a
light on the Gaussian eyepiece. (Check that the light hole is
still open). With no illumination on the prism
except that from the telescope, next rotate the prism table back and forth
a few degrees until maximum
reflected light appears in the eyepiece.
Clamp the table in this position, and
then focus the telescope (by turning the focus ring) until the reflected
image of the cross hairs appears and shows no parallax with respect to the
cross hairs. The telescope is now focused for parallel rays. Fine
adjustments of the prism table leveling screws may help
the images coincide and thus set telescope axis accurately
to the reflecting surface. See Part VI, Sec. 3.
- 2.
- To find the prism angle set the telescope
to first one
prism face and then the other. The angle between these two positions is the
supplement of the prism angle. See Part IIA for detail about angle
readings.
PART III: INDEX OF REFRACTION
- Introduction: When the path thru a prism is symmetric, the deviation is
a minimum. At this angle of minimum deviation the index of
refraction,
, is
where = angle of prism
= angle of minimum deviation
= refractive index of the prism. |
![\includegraphics[height=1.8in]{figs/l5-04a.eps}](img408.png) |
Devise your own methodology and
determine the prism's refractive index for one or more lines
of the Hg spectrum.
PART IV: CALIBRATION OF PRISM SPECTROSCOPE
- Set the prism for minimum deviation for a green line in the Hg
spectrum.
- Determine the angle readings for the yellow, green,
blue-violet, and deep-violet Hg lines. Repeat this for the red line from a
hydrogen discharge.
- Plot telescope angle vs accepted
's for these lines.
- For the same setting of the prism table, find the angles
for the blue-green and two
violet lines of the hydrogen spectrum.
- Use your calibration curve to determine
of the
blue-green and violet lines in the atomic hydrogen spectrum. Compare
results to accepted values.
OPTIONAL: The calibration curve is very non-linear. Since a more linear
plot facilitates interpolation of an unknown
, try plotting deviation
vs 1/
.
TABLE OF WAVELENGTHS:
Mercury lines |
Hydrogen lines |
|
(in nm) |
|
(l/ ) |
|
(in nm) |
(1/ ) |
Yellow |
579.0
(unresolved) |
|
(nm) |
|
|
(nm) |
Yellow |
577.0
Avg. = 578 |
|
2.993 |
Red |
656.3 |
2.322 |
Green |
546.1 |
|
3.353 |
Blue-green |
486.1 |
4.232 |
Blue |
496.0 |
|
4.065 |
|
|
|
|
49l.6 |
|
4.138 |
Violet |
434.0 |
5.309 |
Blue |
435.8 |
|
5.265 |
|
|
|
-violet |
434.7 |
|
|
Deep Violet |
410.2 |
5.943 |
Deep- |
433.9 |
|
|
|
|
|
violet |
407.8 |
|
6.013 |
NOTE: implies |
|
|
404.7 |
|
6.106 |
usually faint |
|
PART V: DIFFRACTION GRATING
CAUTION: Do not touch grating surface under any circumstance! |
- Grating constant (number of lines per cm): The
lines/cm marked on these replica gratings is only
approximate because the plastic often changes dimensions when
it is stripped from the master grating. To achieve quantitative
results calibration of the grating constant is desireable and can
be performed using the known
, 546.1 nm, of the Hg green line.
CALIBRATION PROCEDURE:
- a)
- Align telescope with collimator so the slit image falls exactly on the
vertical cross hair.
- b)
- Mount grating on the dowel pins in the prism table. Adjust and
clamp the table so the grating is approximately
to telescope-collimator axis.
- c)
- Then adjust the table so that the front grating face is accurately
to telescope. Use the Gaussian eyepiece (see Part II - B)
for the final adjustment: the
reflected image of the cross hairs should fall on the cross hairs. (Two
reflected images can result if the sides of the grating glass
are not parallel. If so, test which image comes from the grating
side by wetting the other side.) Clamp
prism table in this position and record telescope direction.
- d)
- Next set
telescope on the first and second orders of the Hg green line. Use both
angles. The
angles should of course agree. Use them and
to calculate the grating spacing d from
- Measurement of the wavelengths in the Balmer series in hydrogen:
- a)
- Use the calibrated grating spectroscope to measure
for
three or four atomic hydrogen spectral lines.
- b)
- Compare these
's to those from the Balmer formula
or calculate the frequencies of these lines and compare to
- c)
- Use Planck's relation,
, to calculate the energies in
eV of the photons for each frequency and, therefore, the energy change in
the hydrogen atom associated with each frequency.
- d)
- Assuming that each observed
leaves hydrogen in the
state, show the observed transitions on a hydrogen
energy level diagram (drawn to scale; consult textbook).
PART VI: ADJUSTMENT OF A SPECTROMETER
NOTE: Not to be performed without permission of the instructor |
-
- 1.
- Use Gaussian eyepiece and prism face method
to focus telescope for parallel
rays. If necessary adjust prism table to make reflected image of the cross
hairs coincide with the cross hairs.
- 2.
- Remove prism from the table and align telescope and
collimator approximately. Focus the collimator until there is no parallax
between the image of the slit and the cross hairs. Adjust the ring on the
collimator focusing sleeve so that the collimator is in focus with the ``V'' in
the slot. Clamp the telescope in line with the collimator so that the image of
the slit (vertical) falls on the vertical cross hairs. Now rotate the slit
through 90
as permitted by the ``V'' projection and slot. Adjust the level of
the telescope so that the image of the slit falls on the horizontal cross hair.
The axes of telescope and collimator are now in line but not necessarily
perpendicular to the axis of rotation of the instrument.
- 3.
-
Bring the telescope to within
of the collimator and use a
face of the prism to reflect light from the collimator down the telescope
tube. Adjust the prism face so that the center of the slit image falls on
the intersection of the cross hairs. The prism face is now parallel to the
instrument axis.
- 4.
- Set the telescope
to the adjusted prism face (use Gaussian eyepiece method).
Return to part 3 setup and adjust
the collimator so that the center of the slit image falls on the
cross hair's intersection. Telescope and collimator are now properly
focused and are
to the instrument axis.
- 5.
- To adjust the second prism face
to the axis of the telescope (again use the Gaussian eyepiece method).
Turn only the proper
screw on the prism table: otherwise one disturbs the adjustment of the first
face. Recheck the first face and, if necessary, readjust it. Continue
this process until both faces are parallel to the axis of the instrument.
Turn collimator slit to vertical position. The spectrometer should
now be in adjustment.
Next: L-8: Polarization
Up: Light
Previous: L-3: Optical Instruments
  Contents
Physics Laboratory
2001-08-28